Sensor Material Specifications
|
|
Housing (Size 10 / 15 / 20)
|
PTFE
|
Wetted Parts, Body / Tube
|
PFA / PFA
|
End Cap (Size 04 / 06)
|
PP
|
Cable Fitting
|
PP
|
Cable Jacket
|
PVC
|
Sensor Specifications
|
|
Speed of Sound in Fluid
|
1000 to 2200 m/s
|
Fluid Temperature*
|
10 C to 60 C
|
Fluid Pressure
|
0 to 70 psig
|
Fluid Kinematic Viscosity*
|
0.5 x 10-6 m2/s to 4.5 x 10-5 m2/s
|
Enclosure Classification
|
IP65
|
Signal Converter Specifications
|
USC-B01 (without display) & USC-D01 (with display)
|
Input Power
|
24 Vdc ± 10%
|
Power Consumption
|
3.3 W Max. Approx @ 140 mA (Continuous) Max. 200 mA, start-up current
|
Ambient Power
|
0°C to 50°C / 30 ~ 80 RH
|
Installation
|
DIN Rail Mounting
|
Enclosure Class
|
IP 20 (Indoor use)
|
Materials
|
USC-B01 - Polycarbonate
|
Weight
|
USC-B01 - 225 grams / USC-D01 - 530 grams
|
EMC
|
EN61326
|
Sensor Cable
|
Shielded twisted pair
|
Electrical Connection
|
Pluggable screw terminals, accepts 28 to 16 AWG wire
|
Analog Output
|
Current O/p: Active: 4 to 20 mA (Max. Load < 900 ohms) Passive: 4 to 20 mA (Max. Load < 1200 ohms @ 30 V max.) Voltage O/p: 0 to 10 Vdc (min. impedance > 10 K)
|
Pulse Output
|
Open Collector (sinking)/Max. 30 Vdc, 200 mA max. Pulse (Hz) - 10 KHz max., Configurable Pulse (Volume) - Pulse weight from 0.1 mL to 1000000 Gallons, Configurable
|
Digital Output
|
Open Collector / Max. 30 Vdc, 200 mA max. two digital Output configurable for Low flow rate/High flow rate/Totalizer greater than/Totalizer less than/Signal strength less/Positive totalizer pulse LED Alarm indication
|
Digital Input
|
Zeroing - Momentarily pull-up to 24Vdc for activation Totalizer Reset - Contact closure or Vdc pull-up (5 Vdc max)
|
Low Flow Cut-off
|
Configurable
|
Linearization
|
Max. 14 points (Configurable)
|
Display (optional)
|
LCD: 2 line 16 digits alpha numeric (with backlight) with LED alarm indication
|
Digital Communication
|
RS 485 and RS 232
|